five

A Dual-Laser Sensor Based on Off-Axis Integrated Cavity Output Spectroscopy and Time-Division Multiplexing Method

收藏
科学数据银行2022-11-01 更新2026-04-23 收录
下载链接:
https://www.scidb.cn/detail?dataSetId=719cd26d740343b88fb9e575b305622b
下载链接
链接失效反馈
官方服务:
资源简介:
In this article, a compact dual-laser sensor based on an o-axis integrated-cavity output spectroscopy and time-division multiplexing method is reported. A complete dual-channel optical structure is developed and integrated on an optical cavity, which allows two distributed feedback (DFB) lasers operating at wavelengths of 1603 nm and 1651 nm to measure the concentration of CO2 and CH4, simultaneously. Performances of the dual-laser sensor are experimentally evaluated by using standard air (with a mixture of CO2 and CH4). The limit of detection (LoD) is 0.271 ppm and 1.743 ppb at a 20 s for CO2 and CH4, respectively, and the noise equivalent absorption sensitivities are 2.68 X 10-10 cm-1 Hz-1/2 and 3.88 X 10-10 cm-1 Hz-1/2, respectively. Together with a commercial instrument, the dual-laser sensor is used to measure CO2 and CH4 concentration over 120 h and verify the regular operation of the sensor for the detection of ambient air. Furthermore, a first-order exponential moving average algorithm is implemented as an effective digital filtering method to estimate the gas concentration.
提供机构:
Kunyang Wang; Anhui Institute of Optics and Fine Mechanics
创建时间:
2022-10-31
5,000+
优质数据集
54 个
任务类型
进入经典数据集
二维码
社区交流群

面向社区/商业的数据集话题

二维码
科研交流群

面向高校/科研机构的开源数据集话题

数据驱动未来

携手共赢发展

商业合作