five

LA-ICP-MS analysis of trace element characteristics and geological implications in sphalerites from Fozichong lead-zinc ore field, Guangxi

收藏
DataCite Commons2025-04-27 更新2025-04-16 收录
下载链接:
https://www.scidb.cn/detail?dataSetId=404c2e3d5b2a4c8ca5c82cfdec306230
下载链接
链接失效反馈
官方服务:
资源简介:
LA-ICP-MS trace element test was completed in Nanjing jupu detection technology co., ltd. For sulfide, USGS MASS-1 and GSE-1G of basaltic molten glass are used to calibrate the sulfide element content. The 193 nm ArF excimer laser ablation system is manufactured by Teledyne Cetac Technologies, and the model is Analyte Excite. The quadrupole inductively coupled plasma mass spectrometer (ICP-MS) is manufactured by Agilent Technologies and the model is Agilent 7700x. The deep ultraviolet (UV) beam generated by excimer laser generator is focused on the mineral surface through homogenization light path, with energy density of 6.06 J/cm2, beam spot diameter of 40 μm, frequency of 6 Hz, and total erosion of 40 s. The original test data is processed offline by ICP MS Data Cal software, and the element content is quantitatively calculated by "no internal standard-matrix normalization method". Before the formal test, all the points to be tested are pre-eroded by large beam spots to eliminate the potential pollution on the mineral surface. In the test stage, the gas background was collected for 15 s, then eroded for 40 s with a beam spot of 40 μm and a frequency of 5 Hz. The aerosol was sent out of the erosion pool by helium, mixed with Ar, and then entered the ICP-MS to complete the test.
提供机构:
Science Data Bank
创建时间:
2023-11-10
二维码
社区交流群
二维码
科研交流群
商业服务