Pre-contoured superior locking plates offer poor bone fit for midshaft clavicle fracture fixation: cadaveric analysis of 4 commercially available systems
收藏DataCite Commons2022-06-02 更新2024-07-29 收录
下载链接:
https://scielo.figshare.com/articles/dataset/Pre-contoured_superior_locking_plates_offer_poor_bone_fit_for_midshaft_clavicle_fracture_fixation_cadaveric_analysis_of_4_commercially_available_systems/19962127
下载链接
链接失效反馈官方服务:
资源简介:
ABSTRACT Objective: The goal of this study is to verify how commercially available pre-contoured superior plates fit on clavicle midshaft fractures. Methods: 100 cadaveric clavicles were evaluated by three distinct observers applying the clavicle congruence score and comparing four different 6 to 8-hole pre-contoured anatomic locking-plate systems. Results: the inter-observer agreement was considered moderate by the percentage agreement and fair by the Fleiss’ Kappa, with no significant differences between evaluations. Only 1 of the 8 plates presented an anatomic fit greater than 70%. Long plates (8 holes) presented a poor fit compared to short plates (6 or 7 holes). Conclusions: the overall evaluation showed that currently-available pre-contoured superior plate systems provide a poor fit on clavicles for midshaft fracture fixations. Long plates present a worse fit compared to short ones.
提供机构:
SciELO journals
创建时间:
2022-06-02