Comparison of signal, noise, SNR and CNR between the reconstructions FBP/QDS, AIDR 3D mild, AIDR 3D standard and AIDR 3D strong.
收藏NIAID Data Ecosystem2026-03-08 收录
下载链接:
https://figshare.com/articles/dataset/_Comparison_of_signal_noise_SNR_and_CNR_between_the_reconstructions_FBP_QDS_AIDR_3D_mild_AIDR_3D_standard_and_AIDR_3D_strong_/1406108
下载链接
链接失效反馈官方服务:
资源简介:
Values are given in arithmetic mean (SD); reconstruction with filtered back projection/ quantum denoising filtering system (FBP/QDS), adaptive iterative dose reduction (AIDR) 3D mild (MILD), standard (STD) and strong (STR); The values of all measurement points are summarised; First, Repeated Measures ANOVA overall analysis including every measurement point as dependent variable showed p<0.001 (p ANOVA) for signal, noise, SNR and also CNR. Subsequently, t-test was used as single test for the 4 quantitative parameters. Bonferroni correction was automatically performed for the multiple testing with 6 possibilities: p1 (FBP/QDS-AIDR 3D mild), p2 (FBP/QDS-AIDR 3D standard), p3 (FBP/QDS-AIDR 3D strong), p4 (AIDR 3D mild-AIDR 3D standard), p5 (AIDR 3D mild-AIDR 3D strong), p6 (AIDR 3D standard-AIDR 3D strong); signal presents the density in Hounsfield Units; noise is SD of the signal; SNR = signal in the vessel measurement point/noise in the aorta ascendens measurement point; CNR = (signal in the surrounding tissue of the vessel measurement point—signal in the vessel measurement point)/noise in the aorta ascendens measurement point.
Comparison of signal, noise, SNR and CNR between the reconstructions FBP/QDS, AIDR 3D mild, AIDR 3D standard and AIDR 3D strong.