Dataset for "Personalised HTO with patient specific plates has mechanical safety equivalent to generic HTO: findings from a novel case-control in silico clinical trial"
收藏DataCite Commons2024-09-19 更新2025-04-17 收录
下载链接:
https://researchdata.bath.ac.uk/id/eprint/926
下载链接
链接失效反馈官方服务:
资源简介:
To compare the mechanical safety of a new personalised 3D printed high tibial osteotomy (HTO) device with an existing generic device, a novel case-control in silico virtual clinical trial was conducted (ClinicalTrial.gov NCT03419598). Twenty-eight knee osteoarthritis patients underwent CT scanning to create a virtual cohort; the cohort was duplicated to form two arms, Generic and Personalised, on which virtual HTO was performed. Finite element analysis was performed to calculate the stresses in the plates arising from simulated physiological activities at three healing stages, simulating 2, 4 and 12 weeks after surgery.
The dataset includes the demographic information of the subjects, the applied loads, the summarised output data from the finite element models, namely the maximum Von Mises stress (MPa) in the plate and the maximum Von Mises strain (unitless) in the tibial bone adjacent to the screws holding the plate to the tibia.
提供机构:
University of Bath
创建时间:
2021-04-15