Validity of the Apple Watch<sup>®</sup> for monitoring push counts in people using manual wheelchairs

收藏
NIAID Data Ecosystem2026-03-12 收录
官方服务:

资源简介:

Objective: A recent Apple Watch® activity-monitoring innovation permits manual wheelchair users to monitor daily push counts. This study evaluated the validity of the Apple Watch® push count estimate. Design: Criterion validity. Setting: Southern Finland and Southeast Queensland, Australia. Participants: Twenty-six manual wheelchair users from Finland and Australia were filmed completing a standardized battery of activities while wearing the Apple Watch® (dominant wrist). Outcome Measures: Wheelchair pushes as determined by the Apple Watch® were compared to directly observed pushes. Results: Agreement between Apple Watch® push counts and directly observed pushes was evaluated using Intraclass correlation coefficients (ICC), Pearson correlations and Bland-Altman analyses. Apple Watch® pushes and directly observed push counts were strongly correlated (ICC = 0.77, P < 0.01) (r = 0.84, P < 0.01). Bland Altman plots indicated that the Apple Watch® underestimated push counts (M = −103; 95% ULoA = 217; LLoA = −423 pushes). Mean absolute percentage error was 13.5% which is comparable to studies evaluating agreement between pedometer-based step counts and directly observed steps. Conclusion: Apple Watch® push-count estimates are acceptable for personal, self-monitoring purposes and for research entailing group-level analyses, but less acceptable where accurate push-count measures for an individual is required.