Ambient ionization techniques enable mass spectrometry (MS) to expand into broader experimental contexts, although it is increasingly clear that results are influenced by the environmental conditions at the site of sampling. Desorption electrospray ionization (DESI), in particular, is affected by variations in relative humidity (RH) levels. Here we describe the design, development, and construction of an environmental control chassis that can actively modulate RH within ± 3% of user-defined set points across a broad humidity range (15%–70% RH). Preliminary characterization demonstrated differential analyte responses across a range of set points, with observed enhancement of leucine-enkephalin, sulfadimethoxine (negative mode), and maltose (positive mode) in response to increased humidity. The measurable differences in analyte signals across discrete humidity set points underscore the importance of environmental control in ambient ionization strategies. The humidity control system outlined here can be translated to other DESI platforms, with construction information provided herein.