Limitations of Vitek GPS-418 Cards in Exact Detection of Vancomycin-Resistant Enterococci with the vanB Genotype
收藏PubMed Central2026-05-16 收录
下载链接:
https://pmc.ncbi.nlm.nih.gov/articles/PMC86825/
下载链接
链接失效反馈官方服务:
资源简介:
The susceptibilities of 20 strains of vancomycin-resistant enterococci (VRE) with the vanB genotype obtained by using Vitek GPS-418 cards were compared with those obtained by the broth dilution method of the National Committee for Clinical Laboratory Standards (NCCLS) (approved standard M7-A4) and with those obtained by the agar screen method using bile esculin azide agar containing 6 μg of vancomycin per ml. Although both the broth dilution and agar screen methods disclosed no discordance, Vitek GPS-418 cards yielded a very major error compared with the results obtained by the reference broth dilution method of the NCCLS. Vitek GPS-418 cards were therefore found to have considerable room for improvement for the accurate detection of vanB VRE strains.