GC electrode functionalization, DEIS studies
收藏Most Wiedzy Open Research Data Catalog2026-04-17 收录
下载链接:
https://mostwiedzy.pl/en/open-research-data/gc-electrode-functionalization-deis-studies,31710080737310-0
下载链接
链接失效反馈官方服务:
资源简介:
This dataset contains experimental results from electrochemical studies of aptamer-based biosensors composed of self-organised AuNC at GC electrode for the rapid detection of E. coli in real human urine samples through polymerase RNA. The study discusses the electrode in each functionalisation step by attaching a non-coding strand of the oligonucleotide and attaching the complementary strand in the hybridisation process. The dataset includes functionalisation steps that lead to an increase in the thickness of the layer functionalised at the electrode surface, which successively increases the charge transfer resistance, RCT and differences in the charge transfer kinetics process. All electrochemical measurements were performed on an Autolab M204 potentiostat (Metrohm) using a three-electrode system. CV and EIS measurements were conducted in a solution consisting of equimolar amounts of 1 mM K3[Fe(CN)6 ] and 1mM [Fe(CN)6] dissolved in 0.01 M PBS (pH 7.4). The impedance monitoring (DEIS) was performed using an AC signal consisting of a package of 21 elementary sinusoids, ranging from 3 to 4500 Hz, superimposed on a DC voltammetry scan, from − 0.35 V to 0.60 V, scan rate 2 mV/s. The DEIS measurements were used to perform the multiparametric impedance discriminant analysis (MIDA).
提供机构:
Adrian Koterwa