遇见数据集

Supplementary Material for: Which Nebulizer Position Should Be Avoided? An Extended Study of Aerosol Delivery and Ventilator Performance during Noninvasive Positive Pressure Ventilation

收藏
Figshare2017-12-06 更新2026-04-29 收录
官方服务:

资源简介:

Background: Research on the effect of nebulizer location on aerosol delivery during noninvasive ventilation has reached inconsistent conclusions. Objective: To investigate the effects of nebulizer position on aerosol delivery efficiency and ventilator performance during noninvasive ventilation. Methods: The Active Servo Lung 5000 respiratory simulation system (ASL5000) was used to simulate a COPD patient. The noninvasive ventilator was set to the spontaneous breathing mode. Six nebulizer positions, 2 exhalation valve types (single-arch exhalation port and whisper swivel), 4 combinations of inspiratory and expiratory pressure, and 2 respiratory rates were used. Results: Significant differences between nebulizer positions existed in aerosol delivery (p p Conclusions: The closer the nebulizer was to the exhalation valves or ventilator, the lower the aerosol delivery efficiency. Nebulizer position had little clinically significant effect on ventilator performance.

创建时间:
2017-12-06
二维码
社区交流群
二维码
科研交流群
商业服务