five

Optimized effective dose from imaging system in radiotherapy

收藏
DataCite Commons2024-05-10 更新2024-07-13 收录
下载链接:
https://researchdata.up.ac.za/articles/dataset/Optimized_effective_dose_from_imaging_system_in_radiotherapy/25670772/1
下载链接
链接失效反馈
官方服务:
资源简介:
The study was conducted at Steve Biko Academic Hospital in the Radiotherapy department. Computed tomography dose index imaging dose measurements were performed using a head and body phantom with slots to insert a calibrated pencil ionisation chamber connected to an electrometer used to measure doses. The computed tomography dose index dose was measured for different cancer treatment setup verification imaging protocols, which included head and neck, pelvis and prostate protocols. Measurements were conducted for 90, 100, 110, 120 and 140 kV, respectively. The measured CBCT dose for each imaging protocol was converted into the effective dose by applying conversion factors. These factors were obtained from the American Association of Physicists in Medicine (AAPM) report no 96. The results were examined, using the mean, median, standard deviation, and confidence intervals, as well as specialized tests such as the Mann-Whitney U test and the intra-class correlation coefficient (ICC).
提供机构:
University of Pretoria
创建时间:
2024-05-10
5,000+
优质数据集
54 个
任务类型
进入经典数据集
二维码
社区交流群

面向社区/商业的数据集话题

二维码
科研交流群

面向高校/科研机构的开源数据集话题

数据驱动未来

携手共赢发展

商业合作