five

Photoacoustic tomography versus cone-beam computed tomography versus micro-computed tomography: Accuracy of 3D reconstructions of human teeth

收藏
Mendeley Data2024-04-13 更新2024-06-28 收录
下载链接:
https://datadryad.org/stash/dataset/doi:10.5061/dryad.cc2fqz699
下载链接
链接失效反馈
官方服务:
资源简介:
Objectives: In this in-vitro study, teeth were imaged using photoacoustic tomography (PAT), cone-beam computed tomography (CBCT), and micro-computed tomography (µ-CT). The study had the aim: to identify the best wavelength for PAT images to determine the accuracy of the three imaging methods and to determine whether PAT images of teeth can achieve acceptable reconstruction quality. Methods: Nineteen human mandibular single-rooted incisors were extracted from patients with trauma or periodontitis. To determine the best wavelength for acquiring photoacoustic images, all 19 teeth were scanned in vitro with PAT, using different laser wavelengths between 680 and 960 nm. The images were analyzed using image analysis software. To assess the accuracy of PAT and compare it with the accuracy of CBCT, each tooth was also scanned in vitro using CBCT and the reference standard technique of µ-CT. Subsequently, three different three-dimensional models, one for each imaging technique, were created for each tooth. Finally, the three different three-dimensional models acquired for the same tooth were matched and analyzed regarding volume and surface. Results: The highest quality tooth images were achieved using the 680 nm wavelength, which showed the best contrast ratio. The full geometry of the dental root (µ-CT compared with PAT) could be visualized with relative standard deviations of 0.12 mm for the surface and -7.33 mm3 for the volume (n=19). The full geometry of the dental root (µ-CT compared with CBCT) could be visualized with relative standard deviations of 0.06 mm for the surface and -14.56 mm3 for the volume (n=19). The difference between the PAT–µ-CT group and CBCT–µ-CT group regarding the total average of the root surface area was not significant (p > 0.06). Conclusion: Images, which were acquired using PAT at 680nm showed the best contrast ratio, enabling the identification of dentin, cementum and the dental pulp. No significant differences were found between the PAT–µ-CT group and CBCT–µ-CT group regarding the total average of the RSA and the total volume. Thus, three-dimensional reconstructions based on in-vitro PAT are already of acceptable reconstruction quality.
创建时间:
2023-06-28
二维码
社区交流群
二维码
科研交流群
商业服务