five

Updating the ELISA standard curve fitting process to reduce uncertainty in estimated microcystin concentrations MethodsX

收藏
NOAA Institutional Repository2022-12-21 更新2026-04-25 收录
下载链接:
https://doi.org/https://doi.org/10.1016/j.mex.2018.03.011
下载链接
链接失效反馈
官方服务:
资源简介:
This study is aimed at exploring the optimal ELISA standard curve fitting process for reducing measurement uncertainty. Using an ELISA kit for measuring cyanobacterial toxin (microcystin), we show that uncertainty associated with the estimated microcystin concentrations can be reduced by defining the standard curve as a four-parameter logistic function on the natural log concentration scale, instead of the current approach of defining the curve on the concentration scale. The model comparison method is outlined in this paper, allowing it to be transferable to test different statistical models for other ELISA test kits. Grant no. NA12OAR4320071
提供机构:
NOAA
创建时间:
2022-12-21
二维码
社区交流群
二维码
科研交流群
商业服务