45° dataset from cubic insulin recorded at the I23 beamline of Diamond Light Source
收藏资源简介:
450 miniCBF X-ray diffraction images from a 45° rotation scan of a cubic insulin crystal recorded at the I23 beamline of Diamond Light Source as part of commissioning tests. This dataset is provided as a small, but complete, diffraction dataset from I23, which has the unique DECTRIS P12M detector. A simple script for processing the data using DIALS 3.10 is as follows: dials.import first450/data_1_*.cbf\ fast_slow_beam_centre=1080,57,12 dials.find_spots imported.expt d_max=75 dials.index imported.expt strong.refl dials.refine indexed.{expt,refl} dials.integrate refined.{expt,refl} dials.symmetry integrated.{expt,refl} dials.scale symmetrized.{expt,refl} The initial beam centre read from the image headers is not correct, so the corrected value is enforced on input in the above script.