Modeling and simulation of cochlear perimodiolar electrode based on composite spring-mass model
收藏DataCite Commons2024-02-16 更新2024-07-29 收录
下载链接:
https://tandf.figshare.com/articles/dataset/Modeling_and_simulation_of_cochlear_perimodiolar_electrode_based_on_composite_spring-mass_model/14987069/2
下载链接
链接失效反馈官方服务:
资源简介:
This paper proposes, a method for the physical modeling of the perimodiolar electrode, particularly for the process of recovering its preset shape with the guide wire drawn out, based on the composite spring-mass model by employing the virtual-volumetric spring inspired from the traditional spring-mass model. Simulation experiments of modeling and virtual insertion of perimodiolar electrode were carried out. The results indicated that the mean and standard deviation of the difference between the local deformation angles of the simulated and measured sets of mass points, (1, 2, 3), (2, 3, 4), …, (13, 14, 15), were 6.34° and 5.98°, respectively. Additionally, the physical model of the perimodiolar electrode can reflect the overall morphological changes of the real perimodiolar electrode.
提供机构:
Taylor & Francis
创建时间:
2022-02-17