EMUE-RMG-1-GasFlowCalibration
收藏资源简介:
Flow meter calibration using the master meter method This example demonstrates the calibration of a gas flow measuring instrument by the so-called “master meter” method, i.e. by comparing the measured flow on a master meter (reference standard) and the measured flow on the device under test. Files contained in dataset are: Flow_meter_calibration_using_the_master_meter_method.pdf RMG11_ExampleSAGAS.tex: LaTeX file to be compiled in order to produce Flow_meter_calibration_using_the_master_meter_method.pdf, EMPIR.jpg: figure file, EMUELogo-03.jpg: figure file, SAGAS1.png: figure file, SAGASGUMandMCM.jpg: figure file, RMG11.bib: bibliography file, GUM_SAGAS.mat: MATLAB function that calculates measurement uncertainty by following the GUM principle, GUMSupp1_SAGAS.mat: MATLAB function that calculates measurement uncertainty by following GUM Supplement 1 principle (Monte Carlo method), InputData.mat: MATLAB function for input data, MainScript.mat: Model developed in MATLAB that calculates estimated gas flow error and the associated measurement uncertainty (GUM and Monte Carlo method). This file that uses MATLAB functions (developed for this example): SetFormattingSymbols, InputData, GUM_SAGAS, GUMSupp1_SAGAS, Results, PlotGUMAndMCM_SAGAS, PlotGUMAndMCM_SAGAS.mat: MATLAB function for plotting the results, Results.mat: MATLAB function for displaying the results, SetFormatingSymbols.mat: MATLAB function for setting formatting symbols.