Upgrade of a Scanning Confocal Microscope to a Single-Beam Path STED Microscope
收藏Figshare2016-01-15 更新2026-04-29 收录
下载链接:
https://figshare.com/articles/dataset/_Upgrade_of_a_Scanning_Confocal_Microscope_to_a_Single_Beam_Path_STED_Microscope_/1455750
下载链接
链接失效反馈官方服务:
资源简介:
By overcoming the diffraction limit in light microscopy, super-resolution techniques, such as stimulated emission depletion (STED) microscopy, are experiencing an increasing impact on life sciences. High costs and technically demanding setups, however, may still hinder a wider distribution of this innovation in biomedical research laboratories. As far-field microscopy is the most widely employed microscopy modality in the life sciences, upgrading already existing systems seems to be an attractive option for achieving diffraction-unlimited fluorescence microscopy in a cost-effective manner. Here, we demonstrate the successful upgrade of a commercial time-resolved confocal fluorescence microscope to an easy-to-align STED microscope in the single-beam path layout, previously proposed as “easy-STED”, achieving lateral resolution in vitro as well as through immunofluorescence recordings of microtubules in a complex epithelial tissue. Here, we applied a recently proposed deconvolution approach and showed that images obtained from time-gated pulsed STED microscopy may benefit concerning the signal-to-background ratio, from the joint deconvolution of sub-images with different spatial information which were extracted from offline time gating.
创建时间:
2016-01-15