Calibrating Single-Ended Fiber-Optic Raman Spectra Distributed Temperature Sensing Data
收藏DataONE2021-12-05 更新2024-06-08 收录
下载链接:
https://search.dataone.org/view/sha256:424f04d9bc9f430aa44004ce625de8c937b33c8664ddb362f13111db98a03caf
下载链接
链接失效反馈官方服务:
资源简介:
Hydrologic research is a very demanding application of fiber-optic distributed temperature sensing (DTS) in terms of precision, accuracy and calibration. The physics behind the most frequently used DTS instruments are considered as they apply to four calibration methods for single-ended DTS installations. The new methods presented are more accurate than the instrument-calibrated data, achieving accuracies on the order of tenths of a degree root mean square error (RMSE) and mean bias. Effects of localized non-uniformities that violate the assumptions of single-ended calibration data are explored and quantified. Experimental design considerations such as selection of integration times or selection of the length of the reference sections are discussed, and the impacts of these considerations on calibrated temperatures are explored in two case studies.
Raw project data is available by contacting ctemps@unr.edu