CCS - MV Bluefin 2017 V01 fCO2
收藏DataCite Commons2020-09-20 更新2025-04-09 收录
下载链接:
https://data.csiro.au/collections/#collection/CIcsiro:33400v1
下载链接
链接失效反馈官方服务:
资源简介:
A General Oceanics model 8050 fCO2 analyser was installed in the carbon team’s container and measured seawater from the ship’s auxiliary seawater pump for the duration of the voyage.
Component sensors:
Sensor | Make Model | Calibration
Non-dispersive infra-red CO2 analyser | Licor | LI-7000 | Continuous (see below)
Equilibrator temperature sensor | Hart-Fluke | 1521 | Jan 2017
Survey settings:
The instrument was configured in continuous flow mode with a measurement interval of 30 seconds. Seawater fCO2 measurements are interrupted for about 22 minutes every 4 hours for analysis of calibration gases and atmospheric air.
Calibration Information:
The system ran a set of 4 calibration gases every 4 hours. Calibration gas concentrations are 0.00 ppm, 261.41 ppm, 377.78 ppm and 450.79 ppm. Calibrations are applied in post-processing.
提供机构:
CSIRO
创建时间:
2018-06-20