Calibration and initial validation of a low-cost computer-based screening audiometer coupled to consumer insert phone-earmuff combination for boothless audiometry
收藏Taylor & Francis Group2022-10-03 更新2026-04-16 收录
下载链接:
https://tandf.figshare.com/articles/dataset/Calibration_and_initial_validation_of_a_low-cost_computer-based_screening_audiometer_coupled_to_consumer_insert_phone-earmuff_combination_for_boothless_audiometry/16529477/1
下载链接
链接失效反馈官方服务:
资源简介:
To undertake calibration and preliminary validation of a custom-designed computer-based screening audiometer connected to consumer insert phone-earmuff combination for adult pure tone audiometry. Part 1 involved electroacoustic measurement and biological calibration of a laptop-earphone pair used for the computer-based audiometry (CBA). Part 2 compared CBA thresholds obtained without a sound booth with those measured using the gold-standard clinical audiometry. 17 young normal-hearing volunteers (Part 1) and 43 normal and hearing loss subjects (Part 2) recruited from an audiology clinic via convenience sampling. The transducer-device combination produced outputs suitable for measuring thresholds down to 0 dB HL. Threshold pairs obtained from the CBA and clinical audiometry were highly correlated (Spearman’s correlation coefficient, <i>ρ</i> = 0.92, <i>p</i> < 0.0001) and had a good degree of agreement (mean difference of −1.06 ± 7.63 dB). Also, the CBA showed about 90% sensitivity and 80% specificity for detecting hearing loss based on low (0.5, 1, 2 kHz) and high frequency (4, 8 kHz) pure tone averages of >25 dB HL. The use of a computer-based audiometer application with consumer insert phone-earmuff combination can offer a cost-effective solution for boothless screening audiometry.
提供机构:
Shaharudin, Majdina N.; Seluakumaran, Kumar
创建时间:
2021-08-28



