遇见数据集

Comparison of Bruker Biotyper (R) and Vitek (R) MS matrix-assisted laserdesorption/ionization–time-of-flight massspectrometry platforms for the identificationof filamentous fungi - Supplemental Table 1

收藏
Taylor & Francis Group2024-05-16 更新2026-04-16 收录
官方服务:

资源简介:

Aims: To evaluate the performance of twomatrix-assisted laser desorption/ionization–time-of-flight mass spectrometry platforms to identify molds isolated from clinical specimens. Methods: Fifty mold isolates were analyzed on Bruker Biotyper R ? and Vitek R ? MS platforms. Two Bruker Biotyper extraction protocols were assessed alongside the US FDA-approved extraction protocol for Vitek MS. Results: The Bruker Biotyper modified NIH-developed extraction protocol correctly identified more isolates than Bruker’s protocol (56 vs 33%). For species in the manufacturers’ databases, Vitek MS correctly identified 85% of isolates, with 8% misidentifications. The Bruker Biotyper identified 64%, with no misidentifications. For isolates not in the databases, the Bruker Biotyper did not misidentify any, and VitekMSmisidentified 36%. Conclusion: Both the VitekMS and Bruker Biotyper accurately identified the fungal isolates, however Vitek MS was more likely to misidentify isolates than the Bruker Biotyper.

创建时间:
2023-06-15
二维码
社区交流群
二维码
科研交流群
商业服务