Enhancing Sensitivity of Microflow-Based Bottom-Up Proteomics through Postcolumn Solvent Addition
收藏Figshare2019-05-22 更新2026-04-29 收录
下载链接:
https://figshare.com/articles/dataset/Enhancing_Sensitivity_of_Microflow-Based_Bottom-Up_Proteomics_through_Postcolumn_Solvent_Addition/8223806
下载链接
链接失效反馈官方服务:
资源简介:
The introduction of more sensitive mass spectrometers allows researchers to adapt front-end liquid chromatography (LC) to individual needs for the analysis of complex proteomes. Where absolute sensitivity is not paramount, it is advantageous to switch from a highly sensitive nanoflow-LC setup, the de facto standard platform in mass-spectrometry (MS)-based discovery proteomics, to a more robust, high-throughput-compatible microflow or conventional-flow setup. To enhance the microflow-LC-MS electrospray process of complex proteomic samples, we tested the effects of different solvents, including 2-propanol, methanol, and acetonitrile, pure or as mixture with dimethyl sulfoxide, which were added postcolumn to the eluting sample. Postcolumn addition of organic solvents strongly enhanced the electrospray efficiency in microflow-LC-MS experiments and improved the sensitivity across the entire gradient and for early eluting peptides by up to 10-fold. Postcolumn solvent addition did not negatively affect chromatographic performance and resulted in an overall 28–36% increase in identifications at both the protein and peptide levels. The presented microflow-LC-MS workflow, including postcolumn solvent addition, can be easily adopted on any LC-MS/MS platform.