Small wedge MicroED data of Lysozyme, low-dose MTH1-8DG and high-dose MTH1-8DG
收藏NIAID Data Ecosystem2026-05-02 收录
下载链接:
https://zenodo.org/record/15168592
下载链接
链接失效反馈官方服务:
资源简介:
Small wedge MicroED data of Lysozyme, low-dose MTH1-8DG and high-dose MTH1-8DG.
A. Data information
Dataset 1: Lysozyme
Dataset 2: MTH1 with ligand 8-OXO-2'-DEOXYGUANOSINE-5'-TRIPHOSPHATE (PDB code: 8DG) collected using high fluence.
Dataset 3: MTH1 with ligand 8-OXO-2'-DEOXYGUANOSINE-5'-TRIPHOSPHATE (PDB code: 8DG) collected using low fluence.
B. Data collection
Data was collected with a CMOS CetaD detector installed on a Titan Krios G2 with an autoloader.
Data collection software: EPUD
Cryo-EM infrastructure unit, Scilifelab, Stockholm.
C. Instrument and measurement settings
Instrumental:
Instrument
Titan Krios G3i
Radiation source
FEG
Accelerating voltage
300 kV
Wavelength
0.01968 Å
Probe Type
Nanoprobe
Beam Diameter
1.5 μm
Beam Convergence
Parallel beam
Detector
Ceta-D CMOS detector
Number of pixels in the image
1024 x 1024
Pixel size
5.6 µm x 5.6 µm
Effective Camera Length
3019.12 (lowdose), 3019.12 (highdose) and 1166 (lysozyme)
Hardware Binning
4
Experimental:
Data Type
Electron diffraction data - 3D ED, Microcrystal diffraction (MicroED)
Data collection method
Continuous rotation
Temperature (K) used during data collection
77 K
Number of crystals contributing to the data set
25 (lowdose), 20 (highdose), 61 (lysozyme)
Number of experimental frames
20 frame/data (lowdose), 20 frame/data (highdose), 30 frame/data (lysozyme)
tilt range, exposure time, tilt per frame
lowdose: 20 degree, 1s, 1 degree/frame
highdose: 20 degree, 1s, 1 degree/frame
lysozyme: 15 degree, 1s, 0.5 degree/frame
Software:
Software used for the data collection
EPUD
Software used for processing
AutoLEI with XDS
Authorship and Bibliography
Author(s) of the data
Lei Wang (ESR 10)
Related data
Publication(s)
Files and data formats
Image format
MRC
创建时间:
2025-04-11