Intermediate Megavoltage Photon Beams for Improved Lung Cancer Treatments
收藏Figshare2015-12-18 更新2026-04-29 收录
下载链接:
https://figshare.com/articles/dataset/_Intermediate_Megavoltage_Photon_Beams_for_Improved_Lung_Cancer_Treatments_/1625457
下载链接
链接失效反馈官方服务:
资源简介:
The goal of this study is to evaluate the effects of intermediate megavoltage (3-MV) photon beams on SBRT lung cancer treatments. To start with, a 3-MV virtual beam was commissioned on a commercial treatment planning system based on Monte Carlo simulations. Three optimized plans (6-MV, 3-MV and dual energy of 3- and 6-MV) were generated for 31 lung cancer patients with identical beam configuration and optimization constraints for each patient. Dosimetric metrics were evaluated and compared among the three plans. Overall, planned dose conformity was comparable among three plans for all 31 patients. For 21 thin patients with average short effective path length (50% = 4.68±0.83 and homogeneity index = 1.26±0.06, as compared to 4.95±1.01 and 1.31±0.08 in the 6-MV plans (p 20Gy), 5 Gy (V5Gy), and mean lung dose (MLD) were 7%/20%, 9%/30% and 5%/10%, respectively in the 3-MV plans (p