资源简介:

This dataset contains the mechanical design CAD file for a blood sample container designed to reduce the blood cell sedimentation and plasma separation. By employing a set of partially separated container sections, effective management of the phenomenon was achieved. It was found that the dimensions of the sections need to be sufficiently large and cannot be significantly reduced from this design model, or surface tension will result in excessive material retention/trapping. Prototypes were manufactured from different plastic materials and a protograph of a component made from polycarbonate sheet with CNC milling is included. This project has received funding from the European Union’s Horizon Europe programme under grant agreement No 101057941 (FADEQ). Views and opinions expressed are however those of the author(s) only and do not necessarily reflect those of the European Union or European Innovation Council. Neither the European Union nor the granting authority can be held responsible for them.

创建时间:
2025-10-23