Optimizing parameters for using the parallel auditory brainstem response (pABR) to quickly estimate hearing thresholds
收藏资源简介:
Objectives: Timely assessments are critical to providing early intervention and better hearing and spoken language outcomes for children with hearing loss. To facilitate faster diagnostic hearing assessments in infants, the authors developed the parallel auditory brainstem response (pABR), which presents randomly timed trains of tone pips at five frequencies to each ear simultaneously. The pABR yields high-quality waveforms that are similar to the standard, single-frequency serial ABR but in a fraction of the recording time. While well-documented for standard ABRs, it is yet unknown how presentation rate and level interact to affect responses collected in parallel. Furthermore, the stimuli are yet to be calibrated to perceptual thresholds. Therefore, this study aimed to determine the optimal range of parameters for the pABR and to establish the normative stimulus level correction values for the ABR stimuli. Design: Two experiments were completed, each with a group of 20 adults (18 â ...