Multicolumn Nanoflow Liquid Chromatography System with Accelerated Offline Gradient Generation for Robust and Sensitive Single-Cell Proteome Profiling

收藏
NIAID Data Ecosystem2026-05-02 收录
官方服务:

资源简介:

Peptide separations that combine high sensitivity, robustness, peak capacity and throughput are essential for extending bottom-up proteomics to smaller samples including single cells. To this end, we have developed a multicolumn nanoLC system with accelerated offline gradient generation. One binary pump generates gradients in an accelerated manner to support multiple analytical columns, and a single trap column interfaces with all analytical columns to reduce required maintenance and simplify troubleshooting. A high degree of parallelization is possible, as one sample undergoes separation while the next sample plus its corresponding mobile phase gradient are transferred into the storage loop and a third sample is loaded into a sample loop. Selective offline elution from the trap column into the sample loop prevents salts and hydrophobic species from entering the analytical column, thus greatly enhancing column lifetime and system robustness. With this design, samples can be analyzed as fast as every 20 min at a flow rate of just 40 nL/min with close to 100% MS utilization time, and continuously for as long as several months without column replacement. We utilized the system to analyze the proteomes of single cells from a multiple myeloma cell line upon treatment with the immunomodulatory imide drug lenalidomide.