CUG EPMA_Oxides_METHOD_OlCpx1.0
收藏DataCite Commons2024-06-05 更新2024-07-13 收录
下载链接:
https://ecl.earthchem.org/view.php?id=3279
下载链接
链接失效反馈官方服务:
资源简介:
This method was designed primarily for analysis of Olivine and Clinopyroxene by EPMA (electron probe microanalysis) using a JEOL JXA-8230 electron probe microanalyzer with four wavelength-dispersive spectrometers (WDS) at the Center for Global Tectonics, School of Earth Sciences, China University of Geosciences (Wuhan). 15 kV accelerating voltage, 20 nA probe current, and a 1 micron beam diameter had been used. Dwell times were 10s on element peaks and half that on background locations adjacent to peaks. Raw x-ray intensities were corrected using a ZAF (atomic number, absorption, fluorescence) correction procedure. A series of natural and synthetic SPI standards were utilized and changed based on the analyzing minerals. The following standards were used: Orthoclase (K), Almandine (Al), Diopside (Si, Ca, Mg), Magnetite (Fe), Jadeite (Na), Rhodonite (Mn), Rutile (Ti), Chromite(Cr) and Apatite (P).
提供机构:
Interdisciplinary Earth Data Alliance (IEDA)
创建时间:
2024-06-05