CT machine comparison
收藏Figshare2025-02-17 更新2026-04-28 收录
下载链接:
https://figshare.com/articles/dataset/CT_machine_comparison/28426541
下载链接
链接失效反馈官方服务:
资源简介:
In human CT imaging, opportunistic calculation of bone mineral density (BMD) is used to screen for osteoporosis, where BMD is estimated by Hounsfield Units (HU). There is potential to use HU for estimation of BMD in veterinary imaging. However, HU values can vary between different machines and over time. We undertook a serial measurements study to determine (1) the extent of the difference of the mean HU and noise (HU standard deviation) of eight regions of interest (ROI) of an electron density phantom scanned on two CT machines, a Siemens CT (conventional) machine and the Equina by Asto CT (standing) machine; and (2) to determine if the HU values obtained on each machine differed over time. The mean HU from 30 scans of an electron density phantom were generated from each of the two CT machines (N=60; conventional and standing). Differences between machines were assessed using a linear mixed effects model. The mean HU of each positive value ROI from the conventional machine was lower than for the standing machine (P
创建时间:
2025-02-17



