Development and clinical validation of a non-invasive, beat-to-beat blood pressure monitoring device, compared to invasive blood pressure monitoring during coronary angiography
收藏Figshare2019-03-01 更新2026-04-29 收录
下载链接:
https://figshare.com/articles/dataset/Development_and_clinical_validation_of_a_non-invasive_beat-to-beat_blood_pressure_monitoring_device_compared_to_invasive_blood_pressure_monitoring_during_coronary_angiography/7865699
下载链接
链接失效反馈官方服务:
资源简介:
ABSTRACT Objective To develop and test a beat-to-beat blood pressure monitoring device during coronary angiography, and compare it with invasive blood pressure monitoring. Methods Twenty-eight patients with an indication for hemodynamic study were selected for this investigation, and kept in supine position. Before starting the coronary angiography, they were instructed about the use of the left radial bracelet for beat-to-beat blood pressure monitoring. Results There was a significant difference between the time required for the catheterization laboratory team to acquire the first invasive blood pressure reading and the time to obtain the first beat-to-beat reading (11.1±5.1 and 1.5±1.8, respectively; p
创建时间:
2019-03-01



