Calibration of the Amplitude-Frequency Characteristic of the Schottky detector system at CSRe
收藏科学数据银行2024-11-02 更新2026-04-23 收录
下载链接:
https://www.scidb.cn/detail?dataSetId=d248b6ea6a8b414ab3e7f59d2ec59673
下载链接
链接失效反馈官方服务:
资源简介:
Schottky mass spectrometry utilizing heavy-ion storage rings is a powerful technique for the precise mass and decay half-life measurements of highly charged ions. Because of their intrinsic amplitude-frequency characteristic (AFC), Schottky detector systems exhibit varying sensitivities at different frequencies. Using low-energy electron-cooled stored ions, a new method is developed to calibrate the AFC curve of the Schottky detector system of the Cooler Storage Ring (CSRe) storage ring located in Lanzhou, China. The data recorded the electron-cooled 56Fe26+ ions at an energy of 13.6 MeV/u to measure the AFC curve within the 3 MHz bandwidth around the resonant frequency of the Schottky detector system at CSRe.
提供机构:
Xinliang Yan; Institute of Modern Physics
创建时间:
2024-10-29



