Feasibility of thin-slice abdominal CT in overweight patients using a vendor neutral image-based denoising algorithm: Assessment of image noise, contrast, and quality
收藏资源简介:
The purpose of this study was to investigate whether the novel image-based noise reduction software (NRS) improves image quality, and to assess the feasibility of using this software in combination with hybrid iterative reconstruction (IR) in image quality on thin-slice abdominal CT. In this retrospective study, 54 patients who underwent dynamic liver CT between April and July 2017 and had a body mass index higher than 25 kg/m2 were included. Three image sets of each patient were reconstructed as follows: hybrid IR images with 1-mm slice thickness (group A), hybrid IR images with 5-mm slice thickness (group B), and hybrid IR images with 1-mm slice thickness denoised using NRS (group C). The mean image noise and contrast-to-noise ratio relative to the muscle of the aorta and liver were assessed. Subjective image quality was evaluated by two radiologists for sharpness, noise, contrast, and overall quality using 5-point scales. The mean image noise was significantly lower in group C than in group A (p