Eletrolytic cell to ionic monitoring of wastewater from cassava processing
收藏NIAID Data Ecosystem2026-03-13 收录
数据链接:
官方服务:
资源简介:
Abstract An electric system to measure the ionic solution of the soil solution, containing a electrolytic cell with gold electrodes and bridge resistor was developed and tested with wastewater to following concentration 1.25 x 104 e 1.25 x 10-2 mol L-1 and following temperature 281 K and 297 K. It was adjusted a mathematic model equation according to Nernst to determine the cell potential in function of the ionic activity calculated by generalized Debye-Hückel model activity coefficient. The equipment reproduces the estimated tensions by Nernst equation, corrected by electric current in function of the ionic activity and the solution temperature.
创建时间:
2005-05-01