Data from the July 16 2025 Advanced Accelerator Diagnostics Collaboration run at the SLAC NLCTA
收藏DataCite Commons2026-02-26 更新2026-04-25 收录
下载链接:
https://datadryad.org/dataset/doi:10.5061/dryad.0rxwdbsfh
下载链接
链接失效反馈官方服务:
资源简介:
An integrated detection system, comprised of a thin diamond sensor
incorporated into a compact signal loop, coupled in turn to an
application-specific integrated circuit readout chip, has been tested for
the first time, using picosecond-duration electron pulses from the NLCTA
accelerator at the SLAC National Accelerator Laboratory. For readout frame
rates as high as 120 kHz, the detection system is shown to achieve a
linear dynamic range of greater than 10 bits and a bandwidth of 4-5 GHz,
with no ringing seen on the tail of the output pulse. This dataset
captures results that demonstrate this system to be the best-performing
multi-GHz ionizing particle detection system achieved to date.
提供机构:
Dryad
创建时间:
2026-02-26