Correlation of Clinical Aqueous Flare Grading to Semi-Automated Flare Measurements Using Laser Flare Photometry
收藏DataCite Commons2024-06-04 更新2024-07-29 收录
下载链接:
https://tandf.figshare.com/articles/dataset/Correlation_of_Clinical_Aqueous_Flare_Grading_to_Semi-Automated_Flare_Measurements_Using_Laser_Flare_Photometry/19073874
下载链接
链接失效反馈官方服务:
资源简介:
To compare laser flare photometry (LFP) measurements of aqueous flare with Standardization of Uveitis Nomenclature (SUN) and modified SUN grading. In this prospective study with multicenter design, uveitis patients were classified according to SUN and modified SUN grading scales. LFP was performed with Kowa FM-700 flaremeter. Mean LFP values were compared with SUN and MSUN scores. The study included 475 LFP measurements, of which 216, 48, 150, 31, 28 and 2 had 0, 0.5+, 1+, 1.5+, 2+ and 3+ flare, respectively. LFP values were significantly different between each two consecutive steps for both clinical gradings (all <i>P</i> < .05). Cut-off values for modified SUN grading steps were defined as 5.7, 9.7, 15.7 and 43.2 for 0/0.5+, 0.5+/ 1+, 1+/1.5+ and 1.5/2+ borders of clinical flare, respectively. LFP proves to be an objective measurement in analyzing aqueous flare comparable to both SUN and MSUN clinical grading systems.
提供机构:
Taylor & Francis
创建时间:
2022-01-26