Inadequacies of Sphygmomanometers Used in Emergency Care Services in a Large Capital City in Brazil
收藏Figshare2017-04-01 更新2026-04-29 收录
下载链接:
https://figshare.com/articles/dataset/Inadequacies_of_Sphygmomanometers_Used_in_Emergency_Care_Services_in_a_Large_Capital_City_in_Brazil/6272621
下载链接
链接失效反馈官方服务:
资源简介:
Abstract Background: Hypertension is the main risk factor for cardiovascular diseases. Technical quality of sphygmomanometers is a prerequisite for the correct measurement of arterial pressure. Objectives: To evaluate sphygmomanometers available in emergency services in the city of Belo Horizonte, Brazil. Methods: We performed a cross-sectional, observational, non-interventional study to evaluate characteristics of the sphygmomanometers available in adult emergency services of public and private hospitals in the city of Belo Horizonte, Brazil. We evaluated 337 sphygmomanometers of 25 hospitals - 15 (of 16) public hospitals and 10 (of 12) private hospitals. Results: Twenty-six percent (88/337) of devices were considered inadequate regarding the INMETRO (National Institute of Metrology, Quality and Technology) standards, 39.2% (132/337) for calibration dates, and 54% (188/337) for the mismatching between cuff's and device's brands. In 13 of 25 hospitals (52%), there were no spare cuffs in different sizes for different arm circumferences. Higher adequacy was found for aneroid and mercury sphygmomanometers used in private hospitals (p = 0.038 and p
创建时间:
2017-04-01