five

Characterization data of commercial tracers for magnetic particle imaging

收藏
DataONE2025-03-13 更新2025-04-26 收录
下载链接:
https://search.dataone.org/view/sha256:e5e4b57531f9c1910e757d1f8f3c1f089fa2fd0f7b1eb9d86aad928755eb53ec
下载链接
链接失效反馈
官方服务:
资源简介:
This dataset contains characterization data for various commercially available tracers for magnetic particle imaging (MPI). Currently this includes Synomag-D with plain, NH2, and PEG modified coatings, Perimag, Vivotrax, and Vivotrax Plus. The dataset includes data on particle physical and magnetic properties, like dynamic light scattering, magnetization versus applied field, and zero-field-cooled/field-cooled magnetization measurements, as well as direct evaluations of tracer MPI performance as measured through RELAX and 2D scans on a Magnetic Insight MOMENTUM imager. A table of relevant particle properties is additionally provided as a summary of the processed measurement data for researchers to evaluate directly. All measurements used to generate these values are also provided to allow the evaluation of other analysis methods. Further, this dataset allows for the comparison of several commonly used commercial MPI tracers using a single instrument and consistent procedures. This is us..., Dynamic Light Scattering Hydrodynamic diameter measurements of particles diluted in deionized water were obtained at room temperature using a Brookhaven ZetaPALS as an intensity-weighted lognormal distribution. Iron Concentration Assay Iron concentration of all tracers was determined by a colorimetric assay using 1,10-phenanthroline. 10 µL of commercial tracers were digested in 1 mL of 70% nitric acid overnight at 101 °C. The following day, 10 µL of the digested sample was placed in a quartz 96-well plate and dried on a hot plate at 115 °C. The plate was transferred into an OT2 liquid handling robot (Opentrons, Long Island City, NY, US) where 46 µL of DI water followed by 30 µL of hydroxylamine hydrochloride (8.06 M) was added and allowed to react for 1 hour. 49 µL of sodium acetate (1.22 M) and 75 µL of 1,10-phenanthroline monohydrate (13 mM) were then added to complex with the Fe2+. Absorbance measurements at 508 nm were collected in a SpectraMax M5 microplate reader (Molecular Device..., , # Dataset title: Characterization data of commercial tracers for magnetic particle imaging ### Principal Investigator Contact Information Carlos M. Rinaldi-Ramos Institution: University of Florida Email: [carlos.rinaldi@ufl.edu](mailto:carlos.rinaldi@ufl.edu) ### Alternate Contact Information Name: Eric Daniel Imhoff Institution: University of Florida Email: [eimhof@ufl.edu](mailto:eimhof@ufl.edu) ## Dates of Data Collection Data was collected throughout 2024. ## Funding This work was supported by the National Institute for Biomedical Imaging and Bioengineering under award number R01EB031224. ## Data Sources All data were derived from measurements performed by the authors using the instrumentation described in the methods. ## Recommended Citation for the Data Set Imhoff, Eric D.; Melnyk, Andrii; Rinaldi-Ramos, Carlos M. (2025). Characterization data of commercial tracers for magnetic particle imaging [Dataset]. Dryad. doi: 10.5061/dryad.j3tx95xs0 ## Related Publications...,
创建时间:
2025-03-14
5,000+
优质数据集
54 个
任务类型
进入经典数据集
二维码
社区交流群

面向社区/商业的数据集话题

二维码
科研交流群

面向高校/科研机构的开源数据集话题

数据驱动未来

携手共赢发展

商业合作