Supplementary Material for: Development of a Normative Database for Multifocal Electroretinography in the Context of a Multicenter Clinical Trial
收藏NIAID Data Ecosystem2026-03-10 收录
下载链接:
https://figshare.com/articles/dataset/Supplementary_Material_for_Development_of_a_Normative_Database_for_Multifocal_Electroretinography_in_the_Context_of_a_Multicenter_Clinical_Trial/4519673
下载链接
链接失效反馈官方服务:
资源简介:
Purpose: The aim of this study is to present the
largest normative database using multifocal electroretinography (mfERG)
in the context of a multicenter clinical trial. Methods:
This investigational study included 156 eyes from 78 Caucasian subjects
aged 45-70 years without known ophthalmic disease or diabetes mellitus;
the subjects were recruited from 11 clinical sites in the setting of the
EUROCONDOR project. Standardized mfERG acquisition (103 hexagons per
eye) was established based on the International Society of Clinical
Electrophysiology in Vision. At least one technician per site received
both specialized training and certification. The main variables that
could have influenced the results were considered in the analyses. Results:
The normative database was based on 111 eyes. The overall mean
P1-implicit time (IT) was 33.94 ± 1.70 ms, and the mean P1 amplitude was
30.58 ± 5.20 nV/deg2. Age and gender were independently
related to predictors of P1-IT but not of P1 amplitude. The responses
that were averaged for the 6 rings showed a longer P1-IT time in the
fovea, decreasing progressively to the parafovea and perifovea. By
contrast, P1 amplitude values sharply decreased with retinal
eccentricity. Conclusions: This normative database can be
used as a comparative index of expected normal values in the clinical
setting and for examining the effect of studies testing neuroprotective
agents.
创建时间:
2017-01-05