RAD, limits of agreement analysis, and Clarke error grid analysis between CGMS and CLD values in data sets at different time after CGMS sensor calibration and in different glycemic ranges.
收藏Figshare2015-12-02 更新2026-04-29 收录
下载链接:
https://figshare.com/articles/dataset/_RAD_limits_of_agreement_analysis_and_Clarke_error_grid_analysis_between_CGMS_and_CLD_values_in_data_sets_at_different_time_after_CGMS_sensor_calibration_and_in_different_glycemic_ranges_/674032
下载链接
链接失效反馈官方服务:
资源简介:
Comparisons of RAD and bias were performed by unpaired t-test. Difference in percentage of matched points in Clarke error grid zone A was analyzed by χ2 test. RAD was calculated as absolute difference between time-matched CGMS and CLD value divided by CLD value, multiplied by 100. Bias was defined as the mean of the difference between time-matched CGMS and CLD values (CGMS minus CLD). Euglycemic level was defined as CLD serum glucose concentration of 3.9 to 10.0 mmol/L, and hyperglycemic level was defined as CLD serum glucose concentration greater than 10.0 mmol/L. Upper and lower limit of agreement were defined as bias±1.96 SD of the mean bias. CGMS = continuous glucose monitoring system. CLD = central laboratory device. RAD = relative absolute difference.