Whole body T1 mapping of small animals using prospective gating and variable flip angle imaging

收藏
DataCite Commons2020-09-18 更新2025-04-16 收录
下载链接:
http://ora.ox.ac.uk/objects/uuid:28040887-8aac-49f6-9c8f-91ddcb0627a4
下载链接
链接失效反馈
官方服务:
资源简介:
Prospective gating and automatic reacquisition of data corrupted by respiration motion were implemented in variable flip angle (VFA) and actual flip angle imaging (AFI) MRI scans to enable cardio-respiratory synchronised T1 mapping of the whole mouse. Stability tests of cardio-respiratory gating (CR-gating) and respiratory gating (R-gating) with and without reacquisition were compared with un-gated scans in 4 mice. The automatic and immediate reacquisition of data corrupted by respiration motion is observed to properly eliminate respiration motion artefact. CR-gated VFA scans with 16 flip angles and 32 k-lines per cardiac R-wave were acquired with R-gated AFI scans in a total scan time of less than 14 minutes. The VFA data were acquired with a voxel size of 0.075 mm3. T1 was calculated in the whole mouse with a robust and efficient nonlinear least squares fit of data. The standard deviation in the T1 measurement is conservatively estimated to be less than 6.2%. The T1 values measured from VFA scans with 32 k-lines per R-wave are in very good agreement with those measured from VFA scans with 8 k-lines per R-wave, even for myocardium. As such, it is demonstrated that prospective gating and reacquisition enables fast and accurate T1 mapping of small animals.