Measurement of Warfarin in the Oral Fluid of Patients Undergoing Anticoagulant Oral Therapy
收藏NIAID Data Ecosystem2026-03-07 收录
下载链接:
https://figshare.com/articles/dataset/Measurement_of_Warfarin_in_the_Oral_Fluid_of_Patients_Undergoing_Anticoagulant_Oral_Therapy/130970
下载链接
链接失效反馈官方服务:
资源简介:
BackgroundPatients on warfarin therapy undergo invasive and expensive checks for the coagulability of their blood. No information on coagulation levels is currently available between two controls.
MethodologyA method was developed to determine warfarin in oral fluid by HPLC and fluorimetric detection. The chromatographic separation was performed at room temperature on a C-18 reversed-phase column, 65% PBS and 35% methanol mobile phase, flow rate 0.7 mL/min, injection volume 25 µL, excitation wavelength 310 nm, emission wavelength 400 nm.
FindingsThe method was free from interference and matrix effect, linear in the range 0.2–100 ng/mL, with a detection limit of 0.2 ng/mL. Its coefficient of variation was <3% for intra-day measurements and <5% for inter-day measurements. The average concentration of warfarin in the oral fluid of 50 patients was 2.5±1.6 ng/mL (range 0.8–7.6 ng/mL). Dosage was not correlated to INR (r = −0.03, p = 0.85) but positively correlated to warfarin concentration in the oral fluid (r = 0.39, p = 0.006). The correlation between warfarin concentration and pH in the oral fluid (r = 0.37, p = 0.009) confirmed the importance of pH in regulating the drug transfer from blood. A correlation between warfarin concentration in the oral fluid and INR was only found in samples with pH values ≥7.2 (r = 0.84, p = 0.004).
ConclusionsWarfarin diffuses from blood to oral fluid. The method allows to measure its concentration in this matrix and to analyze correlations with INR and other parameters.
创建时间:
2011-12-02