Comparison of Murex Single-Use Diagnostic System with Traditional Enzyme Immunoassay for Detection of Exposure to Human Immunodeficiency Virus
收藏PubMed Central2026-05-16 收录
下载链接:
https://pmc.ncbi.nlm.nih.gov/articles/PMC119866/
下载链接
链接失效反馈官方服务:
资源简介:
Because a retrospective study detected 13 negative Western blots out of 38 single-use diagnostic system (SUDS)-positive cases over a 1-year period, we performed a prospective study to compare the performance of the SUDS test with that of enzyme immunoassay (EIA). Of 888 SUDS-tested sera, 875 (98.4%) were both SUDS and EIA negative and 5 (0.6%) were SUDS, EIA, and Western blot positive. The rate of SUDS-positive samples decreased from 3.16/month in the retrospective study to 1.33/month in the prospective study. The immunoassays had sensitivities and specificities of 100 and 99.7 (SUDS) and 100 and 99.4% (traditional EIA), respectively. In laboratories with experienced personnel, the SUDS test performs as well as the EIA as a screen for infection with the human immunodeficiency virus.
提供机构:
American Society for Microbiology (ASM)