five

Investigation of Chimeric Reads Using the MinION

收藏
NIAID Data Ecosystem2026-03-10 收录
下载链接:
https://www.ncbi.nlm.nih.gov/sra/ERP022770
下载链接
链接失效反馈
官方服务:
资源简介:
Following a nanopore sequencing run of PCR products of three amplicons less than 1kb, an abundance of reads failed QC due to template/complement mismatch. A BLAST search demonstrated that some of the failed reads mapped to two different genes - an unexpected observation, given that PCR was carried out separately for each amplicon.A further investigation was carried out specifically to search for chimeric reads, using separate barcodes for each amplicon and trying two different ligation methods prior to sample loading. Despite the separation of ligation products, chimeric reads formed from different amplicons were still observed in the base-called sequence.The long-read nature of nanopore sequencing presents an effective tool for the discovery and filtering of chimeric reads. We have found that at least 1.7% of reads prepared using the Nanopore LSK002 2D ligation kit include post-amplification chimeric elements. This finding has potential implications for other amplicon sequencing technologies, as the process is unlikely to be specific to the sample preparation used for nanopore sequencing.
创建时间:
2018-02-21
5,000+
优质数据集
54 个
任务类型
进入经典数据集
二维码
社区交流群

面向社区/商业的数据集话题

二维码
科研交流群

面向高校/科研机构的开源数据集话题

数据驱动未来

携手共赢发展

商业合作