Supplementary Material for: Evaluation of MRI safety of Ru-106 Eye applicators
收藏DataCite Commons2025-06-01 更新2025-01-06 收录
下载链接:
https://karger.figshare.com/articles/dataset/Supplementary_Material_for_Evaluation_of_MRI_safety_of_Ru-106_Eye_applicators/27934143/1
下载链接
链接失效反馈官方服务:
资源简介:
Introduction: Ruthenium-106 brachytherapy is a primary treatment for Uveal Melanoma (UM), the most common intraocular malignancy in adults. This study evaluated the safety of Ru-106 applicators at 3 Tesla (T) MRI and their impact on image quality.
Methods: Magnetic attraction and eddy currents were tested on a 20mm diameter Ru-106 applicator using a nylon string and a porcine eye. Safety criteria were defined by ocular oncologists, comparing magnetic field interactions to the forces exerted on the eye during surgery. Five UM patients were scanned at 3T MRI with the applicator in-situ using both conventional anatomical sequences and scans optimized to reduce metal artifacts.
Results: Minimal magnetic interactions was observed. Eddy currents caused slight lagging during fast movements, and temporary detachment of the applicator of the porcine eye in conditions that were considered unrealistic for clinical scans. Significant susceptibility artifacts compromised image quality of the affected eye.
Conclusion: Patients with Ru-106 applicators can be safely used in 3T MRI with some simple precautions. MR-image quality of the eye was poor due to major susceptibility artefacts, however, imaging of extra-ocular anatomy is feasible.
提供机构:
Karger Publishers
创建时间:
2024-11-30



