five

Continuous online detection method of heacy metals in water based on LIBs technology

收藏
科学数据银行2022-10-12 更新2026-04-23 收录
下载链接:
https://www.scidb.cn/detail?dataSetId=d6fa68d37fb2477eb4c5b5e59e56f712
下载链接
链接失效反馈
官方服务:
资源简介:
In order to meet the requirement of in-situ online monitoring of heavy metals in industrial wastewater, a new method based on laser induced breakdown spectroscopy (LIBS) was proposed to continuously monitor heavy metals in water Line detection method The enrichment method of graphite substrate evaporation is adopted, and the enrichment mode of sample solution, quantitative addition mode of sample solution, and evaporation enrichment mode of sample solution are studied, It has realized the rapid automatic detection of heavy metal elements in water, improved the detection efficiency and detection capability The calibration curves of these six elements were established and the corresponding detection limits were obtained The relative standard deviation of spectral intensity is 9.24%, 10.45%, 12.59%, 11.58%, 15.24%, 6.87% respectively, and the detection limits are 0.03, 0.012 0.0047、0.033、0.071、0.0376 mg/L.。