IntroductionSerum glial fibrillary acidic protein (sGFAP) is a promising biomarker, but its quantification mainly relies on SIMOA, a technology not widely available in clinical practice.
ObjectivesTo evaluate the analytical performance of two high-throughput automated platforms—Alinity® i (Abbott) and Lumipulse® G1200 (Fujirebio)—for sGFAP quantification.
MethodsA retrospective longitudinal study included 107 serum samples from 23 MS patients. sGFAP was measured with SIMOA SR-X®, Lumipulse® G1200, and Alinity® i. Data were log-transformed. Agreement was assessed using Pearson correlations, Passing–Bablok regression, Bland–Altman analysis, and Δlog correlations between visits. Longitudinal differences across platforms were tested with a linear mixed-effects model (platform as fixed effect, SIMOA as reference). Moreover, ΔSIMOA was modeled against ΔLumipulse and ΔAlinity, adjusting for ΔEDSS, phenotype, relapses and new MRI lesions.
ResultsPassing–Bablok regression yielded slopes of 0.85 (SIMOA–Lumipulse), 0.81 (SIMOA–Alinity), and 0.95 (Lumipulse–Alinity), with intercepts of −0.32, −0.35, and −0.05. Mean log-biases were −0.622, −0.733, and 0.109. Correlations between log-means and log-differences were r = 0.26 (p = 0.006), 0.44 (p < 0.0001), and 0.15 (p = 0.13). The mixed-effects model showed no significant Δlog differences relative to SIMOA (p > 0.1). When modeling ΔSIMOA, ΔLumipulse was a significant predictor (β = 0.51; p = 0.002), whereas ΔAlinity showed only a trend (β = 0.31; p = 0.051). No clinical covariates were significantly associated.
ConclusionAutomated platforms, particularly Lumipulse, showed strong concordance with SIMOA supporting the role in analytical monitoring.