Calculated biases between SaO<sub>2</sub> (oxyhemoglobin saturation measured by blood gas analyzer [ABL 825]) and SpO<sub>2</sub> (oxyhemoglobin saturation measured by 3 pulse oximeters) among patients under room air condition.
收藏NIAID Data Ecosystem2026-03-08 收录
数据链接:
官方服务:
资源简介:
The calculated mean biases (SpO2—SaO2) measured by 3 pulse oximeters for the 85%–100% saturation range are shown. Data are presented as the mean ± SD and analyzed by one-way analysis of variance followed by Tukey’s test. *p<0.0001 vs. Nihonkohden OLV-3100 and Nellcor N-BS Calculated biases between SaO2 (oxyhemoglobin saturation measured by blood gas analyzer [ABL 825]) and SpO2 (oxyhemoglobin saturation measured by 3 pulse oximeters) among patients under room air condition.