five

Data_Sheet_1_A Redox Cu(II)-Graphene Oxide Modified Screen Printed Carbon Electrode as a Cost-Effective and Versatile Sensing Platform for Electrochemical Label-Free Immunosensor and Non-enzymatic Glucose Sensor.docx

收藏
NIAID Data Ecosystem2026-03-12 收录
下载链接:
https://figshare.com/articles/dataset/Data_Sheet_1_A_Redox_Cu_II_-Graphene_Oxide_Modified_Screen_Printed_Carbon_Electrode_as_a_Cost-Effective_and_Versatile_Sensing_Platform_for_Electrochemical_Label-Free_Immunosensor_and_Non-enzymatic_Glucose_Sensor_docx/14623704
下载链接
链接失效反馈
官方服务:
资源简介:
A novel copper (II) ions [Cu(II)]-graphene oxide (GO) nanocomplex-modified screen-printed carbon electrode (SPCE) is successfully developed as a versatile electrochemical platform for construction of sensors without an additionally external redox probe. A simple strategy to prepare the redox GO-modified SPCE is described. Such redox GO based on adsorbed Cu(II) is prepared by incubation of GO-modified SPCE in the Cu(II) solution. This work demonstrates the fabrications of two kinds of electrochemical sensors, i.e., a new label-free electrochemical immunosensor and non-enzymatic sensor for detections of immunoglobulin G (IgG) and glucose, respectively. Our immunosensor based on square-wave voltammetry (SWV) of the redox GO-modified electrode shows the linearity in a dynamic range of 1.0–500 pg.mL−1 with a limit of detection (LOD) of 0.20 pg.mL−1 for the detection of IgG while non-enzymatic sensor reveals two dynamic ranges of 0.10–1.00 mM (sensitivity = 36.31 μA.mM−1.cm−2) and 1.00–12.50 mM (sensitivity = 3.85 μA.mM−1.cm−2) with a LOD value of 0.12 mM. The novel redox Cu(II)-GO composite electrode is a promising candidate for clinical research and diagnosis.
创建时间:
2021-05-20
二维码
社区交流群
二维码
科研交流群
商业服务