five

Validation, measurement uncertainty estimation and evaluation of UHPLC greenness for simultaneous determination of metoprolol tartrate and hydrochlorothiazide in binary tablet

收藏
DataCite Commons2024-04-09 更新2024-08-19 收录
下载链接:
https://tandf.figshare.com/articles/dataset/Validation_measurement_uncertainty_estimation_and_evaluation_of_UHPLC_greenness_for_simultaneous_determination_of_metoprolol_tartrate_and_hydrochlorothiazide_in_binary_tablet/25569795/1
下载链接
链接失效反馈
官方服务:
资源简介:
A novel green ultra high performance liquid chromatography (UHPLC) method was validated and estimated the measurement uncertainty for simultaneous determination of metoprolol tartrate (MET) and hydrochlorothiazide (HCT) in binary tablet. A Zorbax<sup>®</sup> SB-C18 column with isocratic elution (flow rate 0.9 mL min<sup>−1</sup>) at 25°C was used. Analytes and HCT degradation product were separated in approximately 1 min using acetonitrile:water:triethylamine (17:83:0.2, v/v) as mobile phase. Analytical curves were linear with (<i>R</i><sup>2</sup> &gt; 0.99 for MET and HCT)  with detection limits of 2.42 and 1.05 µg mL<sup>−1</sup> for MET and HCT, respectively.  Precision measurements showed RSD% from 0.39 to 1.2% and method accuracy by recovery tests was 100 ± 2%. Measurement uncertainties using Eurachem procedure were 100.1 ± 2.8% and 100.3 ± 2.4% for MET and HCT, respectively. This UHPLC method was accurate, precise, linear and selective, making it suitable for pharmaceutical quality control. It also proved eco-friendly compared to other reported methods.
提供机构:
Taylor & Francis
创建时间:
2024-04-09
二维码
社区交流群
二维码
科研交流群
商业服务