Maximum stable gain and fitting range of percutaneously applied bone conduction devices
收藏NIAID Data Ecosystem2026-05-10 收录
下载链接:
https://figshare.com/articles/dataset/Maximum_stable_gain_and_fitting_range_of_percutaneously_applied_bone_conduction_devices/30578015
下载链接
链接失效反馈官方服务:
资源简介:
to measure the maximum stable gain (MSG, the feedback limit) and its variability in users with two percutaneous power bone conduction devices (BCDs), to determine an MSG-based fitting range, and to measure the available aided dynamic range.
The MSG was determined with the BCD attached to a participant’s head. The BCD output was measured on a skull simulator. A fitting range was derived using the speech intelligibility index.
42 adult BCD users.
The average and range of the MSG was 13.3 ± 3.2 dB for device 1 and 16.1 ± 5.90 dB for device 2. The fitting range is 35–40 dB HL for a good understanding of soft speech and 45–50 dB HL for moderate-to-sufficient understanding of speech at a normal level in at least 50% of the cases with this hearing loss. The output for common speech levels is not limited by the maximum force output, except at low frequencies.
For users of a percutaneous power BCD, the maximum applicable gain is limited by feedback and is smaller than the full-on gain. Consequently, BCDs enable sufficient speech intelligibility only for mild-to-moderate sensorineural hearing loss. The aided dynamic range is adequate for hearing thresholds within this fitting range.
创建时间:
2025-11-10



